Evaluation of two HbA1c point-of-care analyzers

  • Catalina Sánchez-Mora , Manuel S. Rodríguez-Oliva , Patricia Fernández-Riejos , Joaquín Mateo , Juan Polo-Padillo , Raimundo Goberna and Víctor Sánchez-Margalet EMAIL logo

Abstract

Background: Measurement of HbA1c is the most important parameter to assess glycemic control in diabetic patients. Different point-of-care devices for HbA1c are available. The aim of this study was to evaluate two point-of-care testing (POCT) analyzers (DCA Vantage from Siemens and Afinion from Axis-Shield). We studied the bias and precision as well as interference from carbamylated hemoglobin.

Methods: Bias of the POCT analyzers was obtained by measuring 53 blood samples from diabetic patients with a wide range of HbA1c, 4%–14% (20–130 mmol/mol), and comparing the results with those obtained by the laboratory method: HPLC HA 8160® Menarini. Precision was performed by 20 successive determinations of two samples with low 4.2% (22 mmol/mol) and high 9.5% (80 mmol/mol) HbA1c values. The possible interference from carbamylated hemoglobin was studied using 25 samples from patients with chronic renal failure.

Results: The means of the differences between measurements performed by each POCT analyzer and the laboratory method (95% confidence interval) were: 0.28% (p<0.005) (0.10–0.44) for DCA and 0.27% (p<0.001) (0.19–0.35) for Afinion. Correlation coefficients were: r=0.973 for DCA, and r=0.991 for Afinion. The mean bias observed by using samples from chronic renal failure patients were 0.2 (range –0.4, 0.4) for DCA and 0.2 (–0.2, 0.5) for Afinion. Imprecision results were: CV=3.1% (high HbA1c) and 2.97% (low HbA1c) for DCA, CV=1.95% (high HbA1c) and 2.66% (low HbA1c) for Afinion.

Conclusions: Both POCT analyzers for HbA1c show good correlation with the laboratory method and acceptable precision.


Corresponding author: Víctor Sánchez-Margalet, Clinical Biochemistry Department, Virgen Macarena University Hospital, Av. Dr. Fedriani 3, Seville 41071, Spain

Received: 2010-7-23
Accepted: 2010-11-2
Published Online: 2011-02-17
Published in Print: 2011-04-01

©2011 by Walter de Gruyter Berlin New York

Downloaded on 2.5.2024 from https://www.degruyter.com/document/doi/10.1515/CCLM.2011.101/html
Scroll to top button