The advent of continuous glucose monitoring (CGM) has allowed significant further insights into diabetes control. Initial CGM devices were predominantly clinician-orientated, as recorded data were only accessible retrospectively. Use of these devices to initiate focused therapeutic changes has been studied extensively with some [14], but not all [5, 6] studies reporting improvements in overall glycaemic control.

In contrast, the newest generation of CGM devices are patient-orientated and offer real-time (RT) access to interstitial glucose levels. Studies assessing their use have shown improvement in HbA1c [711], increased time spent in the target glucose range [11, 12] and fewer hypo- and hyperglycaemic excursions [9, 13], suggesting the potential for important clinical benefits.

Patients using insulin pump therapy may gain maximal benefit from real-time continuous glucose monitoring (RT-CGM), as continuous subcutaneous insulin delivery can be altered in response to real-time glucose trends. Use of RT-CGM with insulin pump therapy can be described as sensor-guided pump management. The MiniMed Paradigm REAL-Time system (Medtronic MiniMed, Northridge, CA, USA) integrates RT-CGM with insulin pump technology. Interstitial glucose readings are transmitted and displayed on the device screen every 5 min for up to 72 h; glucose-trend information and programmable hypo- and hyperglycaemic alarms are further features. Of note, RT data recorded and displayed by the system have no inherent influence on insulin delivery; rather, the need for therapeutic changes must be recognised and initiated by the user.

Previous controlled studies of RT-CGM technology [8, 10, 14] have incorporated regular clinician review of the RT data and clinician-led adjustments to insulin delivery, which may have impacted on the glycaemic improvements accrued. Whether individuals with type 1 diabetes can adapt to and employ RT-CGM to improve their own glycaemic control has not previously been tested. The aim of this study, therefore, was to assess the impact of patient-led use of sensor-guided pump management on indices of glycaemic control in adolescents and young adults with type 1 diabetes and compare the impact with that of standard insulin pump therapy.

Methods

An open randomised controlled trial was carried out at five centres across Australia; participants were recruited from two adult and three paediatric tertiary diabetes clinics. The study protocol received institutional Ethics Committee approval at each of the study sites; written consent was obtained from all participants with additional parent/guardian consent for participants aged <18 years.

Inclusion criteria for participation were: age 13.0–40.0 years, type 1 diabetes for >1 year, use of insulin pump therapy including proficiency with use of a bolus-dose calculator for >3 months, HbA1c ≤8.5%, reliably performing self-monitoring of blood glucose (SMBG) at least four times daily, and internet access. Willingness to use the subcutaneous sensor component of the system for at least 70% of the total 3 month study period was a further protocol requirement. Individuals with co-existent medical problems that would interfere with their ability to use the system (e.g. impaired vision), co-existent illness that otherwise predisposes to hypoglycaemia (e.g. adrenal insufficiency) or a history of severe hypoglycaemia while using insulin pump therapy were excluded. Severe hypoglycaemia was defined as an episode of hypoglycaemia resulting in seizure or coma or requiring third-party assistance or the use of glucagon or intravenous glucose for recovery.

Participants were recruited in age- and sex-matched pairs. Each centre generated an alphabetical list of potential participants who were sequentially invited to participate. When an individual gave consent for participation, the next suitable age- and sex-matched person was approached to complete the pair. Each participant was allocated a chronological study number. Once recruited, a pair of participants was entered, in order of study number, into a computer-generated schedule which randomly assigned each of the pair to one of the two study groups. The randomisation schedule was administered centrally; clinicians involved in participant recruitment had no access to the schedule.

Participants in the intervention group received standardised education in use of the integrated sensor-guided pump system and its associated RT features from the same instructor across all sites. The system requires calibration using a capillary blood glucose measurement at 2 and 8 h post insertion and at least 12 hourly thereafter. Hypo- and hyperglycaemic alarms were set to alert the user at sensor glucose levels <4.5 and >12.0 mmol/l, respectively. Participants were also directed to perform confirmatory capillary blood glucose measurements if RT data suggested the need for therapeutic action (e.g. administration of a correction bolus of insulin). RT data were to be uploaded to a web-based management program (Carelink; Medtronic) every 2 weeks. No specific instructive guidelines for interpreting RT data were provided in this study (‘patient-led’ use). Control group participants continued their usual insulin pump therapy and SMBG regimen. Both groups had access to their routine diabetes care team, but no specific additional contact or advice was initiated by the research team over the study period.

Baseline and end-of-study investigations for all participants comprised 6 days of blinded continuous glucose monitoring using the CGMS Gold (Medtronic) and HbA1c measurement. The difference between groups in the proportion of time spent in the target glycaemic range (defined as 4.0–10.0 mmol/l) was the primary outcome of interest. Secondary outcomes included differences in HbA1c, proportion of time spent in the hypoglycaemic (≤3.9 mmol/l) and hyperglycaemic (≥10.1 mmol/l) ranges and glycaemic variability. All HbA1c measurements were performed at a central independent DCCT-accredited laboratory (using a Primus CLC330 Affinity Chromatography HPLC analyser [Primus Diagnostics, Kansas City, MO, USA]). Glycaemic variability was analysed using continuous overlapping net glycaemic action (CONGA), which allows for assessment of intra-day variability at repeated time intervals [15]. Sensor accuracy was assessed using mean absolute relative difference (MARD), calculated as the average of the absolute percentage difference between a CGMS sensor reading and a corresponding time-matched capillary glucose reading.

Use of the real-time sensor component of the system by participants in the intervention group was determined from data uploaded to the web-based care management program. The impact of adherence with protocol-stipulated ≥70% overall RT sensor usage throughout the study period on end-of-study HbA1c was also assessed.

Serious adverse events were reported to the study co-ordinator at each site throughout the study period. Events monitored and recorded included episodes of severe hypoglycaemia or diabetic ketoacidosis, any hospitalisation and device-related complications or malfunction.

This investigator-initiated study was supported by a research grant and provision of equipment from Medtronic Australasia. The manufacturer requested the hypoglycaemic alarm setting of 4.5 mmol/l (to allow safe recognition of hypoglycaemia despite potentially increased sensor error margin at lower glucose levels), but had no other involvement in study design, data analysis or preparation of the manuscript.

Statistical analyses

Our sample size calculation was based on representative CGM data (n = 92) from individuals established on insulin pump therapy at one study centre (Royal Children’s Hospital, Melbourne, Australia). These data indicated that the mean (SD) proportion of time spent in the target glycaemic range of 4.0–10.0 mmol/l was 54% (33%); a sample of 28 in each group was therefore estimated to have 80% power to detect a 25% difference in mean time spent in target range, with a 0.05 two-sided significance level.

Differences in primary and secondary outcome variables between study groups at the end of the study were assessed using: (1) two-sample Student’s t tests (unadjusted estimates); and (2) linear regression analysis with baseline value as covariate (adjusted estimates). Use of the RT sensor component of the system was calculated as a percentage of the total study period for each intervention group participant (number of recorded sensor readings divided by total possible sensor readings for the 90 day study period [288 × 90] ×100). The impact of adherence with protocol-stipulated use ≥70% on study outcomes was assessed using two-sample Student’s t tests. All analyses were intention to treat and performed using Stata 10 (2007; StataCorp LP, College Station, TX, USA). Results of glycaemic outcomes are expressed as mean ± SD.

Results

Demographic data, sensor accuracy and adherence to study protocol

Sixty-two participants took part in the study; 31 were randomised to each study arm. The study flow diagram is depicted in Fig. 1. Study groups had similar characteristics at baseline (Table 1). Seven participants withdrew from the study: five from the intervention group and two from the control group. The reasons for withdrawal amongst intervention group participants were: burden of recurrent alarms (n = 2); significant skin irritation despite use of various different transmitter adhesives (n = 2); and inability to maintain prolonged skin adhesion of the sensor because of perspiration (n = 1). Two control group participants withdrew consent to undergo end-of-study investigations. End-of-study CGM and HbA1c data were therefore available for comparison on 26/31 participants in the intervention group and 29/31 in the control group. All participants who withdrew were adolescents (mean age 15.6 years, mean HbA1c 7.6%).

Fig. 1
figure 1

Flow of participants through the study

Table 1 Baseline characteristics of all participants randomised to each study group

In total, 3,459 paired sensor and blood glucose readings were recorded over the periods of CGM at baseline and the end of the study. Overall mean and median MARD values were 13.3% and 8.6%, respectively. Variation at different glucose ranges was noted, with mean MARD values of 17.5%, 14.5% and 9.0% in the predefined hypoglycaemic, target and hyperglycaemic ranges, respectively.

Continuous RT data were uploaded to the web-based management program during the 3 month period by 25/26 participants in the intervention group. One participant did not upload any RT-CGM data but is included in the intention-to-treat analyses of principal outcomes as end-of-study investigations were completed. Patterns of RT sensor use varied considerably within the intervention group over the study period. Median (range) time spent using the RT sensor component of the system was 62.5% (17.7–93.8%) of the total 90 day period; 11/25 participants adhered to protocol-stipulated sensor use of ≥70%, while 14/25 did not meet this requirement.

Glycaemic outcomes

Study outcomes are displayed in Table 2. As shown, there was no difference between study groups in the primary outcome of proportion of time spent in the target glycaemic range over 6 days of CGM at the end of the study. In addition, no between-group difference was found in any of the outcomes of time spent hypo- or hyperglycaemic or glycaemic variation.

Table 2 Glycaemic outcomes

Despite the lack of difference in CGM-derived outcome variables, a significant effect on end-of-study HbA1c was evident between study groups. End-of-study mean ± SD HbA1c was 7.1 ± 0.8% in the intervention group vs 7.8 ± 0.8% in the control group, compared with baseline levels of 7.3 ± 0.6% and 7.5 ± 0.7%, respectively. When adjusted for baseline values, mean HbA1c was 0.43% lower in the intervention group compared with that of the control group (95% CI 0.19 – 0.75%, p = 0.009; Table 2). Overall, a reduction in HbA1c was achieved by 16/26 participants (64%) who used sensor-guided pump management compared with 5/29 participants (17%) who used standard pump therapy. HbA1c levels of <7% were also achieved more frequently in the intervention group. Similar proportions in the intervention and control groups (eight [31%] and ten [34%] participants, respectively) met this goal at baseline; however, 14/26 (56%) participants in the intervention group had end-of-study HbA1c levels ≤7%, compared with 5/29 (17%) participants in the control group, p = 0.004.

Associations between RT sensor usage and glycaemic outcomes

Subgroup analysis between those intervention group participants who adhered to protocol-stipulated sensor usage ≥70% (n = 11) and those who failed to meet this requirement (n = 14) was performed. End-of-study HbA1c was 6.7% in participants with ≥70% use and 7.4% in those with <70% sensor use; adjusted for baseline values, HbA1c was 0.51% lower in participants who wore the sensor ≥70% of the total study period (95% CI 0.04–0.98%, p = 0.04; Fig. 2a). The individual change in HbA1c for each intervention group participant is shown in Fig. 2b. Only two participants who wore the sensor ≥70% of the total study period did not have improved glycaemic control overall; in each instance, HbA1c was <7% at both baseline and the end of the study.

Fig. 2
figure 2

Effect of frequency of sensor use on HbA1c within the intervention group. a HbA1c grouped according to adherence with protocol-stipulated sensor use ≥70%: baseline results shown as light grey bars; end-of-study results shown as dark grey bars; p = 0.04 for the difference between subgroups at the end of the study. b HbA1c at baseline and 3 months in each individual participant, plotted according to percentage of sensor use. Cross, HbA1c at baseline; black diamond, final HbA1c in participants with improvement in HbA1c over the study period; white diamond, final HbA1c in participants with deterioration in HbA1c over the study period

Participants with ≥70% sensor use performed a greater number of average SMBG checks per day compared with the less adherent subgroup (5.0/day vs 3.5/day, respectively, p = 0.04). The average number of boluses per day was similar between these subgroups (7.0/day vs 7.8/day, respectively, p = 0.5).

Predictors of more frequent use of RT sensor

The baseline characteristics of participants in the intervention group were analysed to identify potential predictors of increased sensor use (adherence with ≥70% use). No significant effect of age, duration of diabetes or duration of insulin pump therapy was evident: mean age 22.8 years in those with ≥70% sensor use vs 27.2 years in those with <70% use, p = 0.2; mean duration of diabetes was 12.0 years and 12.2 years, respectively, p = 0.96; mean duration of insulin pump therapy use was 1.7 years vs 3.1 years in those with ≥ 70% and <70% sensor use, respectively, p = 0.09.

Adverse events

No episodes of severe hypoglycaemia or diabetic ketoacidosis occurred. One participant in the intervention group was admitted to hospital for treatment of new-onset depression; the study protocol was completed nonetheless. Mechanical problems arose: one participant in the intervention group experienced failure of her study insulin pump device necessitating use of her pre-study model until a replacement pump was delivered the following day. It was also necessary to replace radiofrequency transmitters for four participants in the intervention group for whom these devices developed faults that precluded data storage (the problem was identified as repeated ‘bad sensor’ signals from insertion).

Discussion

This randomised controlled trial (RCT) demonstrates that self-directed use of sensor-guided pump management is associated with modest improvements in glycaemic control when compared with standard pump therapy in individuals with type 1 diabetes. Participants randomised to use sensor-guided management achieved a mean HbA1c level that was 0.43% lower than that of those who continued standard insulin pump therapy. Sensor usage ≥70% positively influenced glycaemic outcomes.

Previous RCTs of this technology have yielded varying results [8, 10, 14, 16]. Two studies have shown improved glycaemic control with use of RT-CGM in patients using either multiple daily injections or insulin pump therapy [8, 10]. Benefits were limited to individuals ≥25 years old in the most recent study [10]. In contrast, RCTs comparing sensor-guided pump management with standard pump therapy alone have found no between-group difference in HbA1c [14, 16]. It is possible that glycaemic benefits seen in participants with suboptimal glycaemic control despite previous intensive management may reflect improved use of pump-related technology (such as bolus-dose calculators) rather than benefits related principally to use of RT-CGM. In addition, most of these studies involved regular clinician review of RT-CGM data and/or clinician-initiated changes to insulin therapy [8, 10, 14]. It has been noted that regular clinician–patient contact is highly labour intensive [17], and there are no agreed guidelines for incorporating RT-CGM data into routine clinical care.

The specific aim of our study, therefore, was to assess whether individuals with relatively well-controlled diabetes on insulin pump therapy could employ RT-CGM to further improve their own glycaemic control. Sensor-guided management was intentionally ‘patient-led’. Participants in both study groups were free to adjust insulin delivery as they deemed necessary; neither group received specific direction on how this should be achieved. The end-of-study difference in HbA1c was due to a combination of improvement in the intervention group and deterioration in the control group. Nonetheless, in light of the absence of a glycaemic threshold for development of microvascular complications in the DCCT [18], the 0.43% difference in HbA1c and the significant increase in numbers achieving target levels of HbA1c <7.0% with use of RT-CGM are clinically meaningful. The non-significant increase in mean HbA1c in the control group over the study period also suggests the relative lack of a ‘Hawthorne-effect’ related to this protocol. Despite the significant end-of-study difference in HbA1c, we were unable to identify differences between groups in any of the CGM-derived glycaemic outcomes. Correlations between CGM data and HbA1c have been described previously [10, 19], but have not been found universally. Despite an end-of-study difference in HbA1c of 0.6%, Deiss et al. [8] found no significant differences in CGM-derived mean glucose, hypo- and hyperglycaemic excursions or area under the curve (D. Deiss, Endokrinologikum, Berlin, Germany, personal communication). These discrepancies may reflect the fact that a 6 day period of continuous glucose monitoring is not consistently representative of longer periods of glycaemic control.

In light of the apparent age effect of sensor-guided management noted in the recent Juvenile Diabetes Research Foundation study [10], we performed post-hoc analyses of age-banded subgroups (13.0–19.0 years and ≥19.0–40.0 years) in our cohort. Interestingly, significant glycaemic benefit was seen only in the 13.0-19.0 years subgroup in our study. After adjusting for baseline value, there was a 0.6% between-group difference in the end-of-study HbA1c in this younger subgroup (95% CI −1.1 to −0.1; p = 0.025); no significant difference was observed in older participants (95% CI −0.7 to 0.2; p = 0.2). CGM-derived study outcomes were not affected by age. These data should be interpreted with caution as analyses were unplanned; however, they suggest that younger participants may indeed profit from this technology. Notwithstanding this, all participants who withdrew from our study were adolescents.

Our study may include unintentional bias. Selection bias was minimised by use of strict recruitment and randomisation procedures, including allocation concealment. The nature of the intervention prevented participant blinding and this may have led to ascertainment bias. Individuals with poor diabetes control were excluded from this study. Thus, our findings may only be applicable to motivated patients who have pre-existing good control on insulin pump therapy.

RT-CGM may be viewed as a behaviour-modification tool; its application requires understanding and interaction from the user and its success is reliant on the manner in which it is employed. The positive association between overall sensor use and glycaemic benefit in our study population has been noted previously [7, 8, 10]. In our cohort, participants with the greatest sensor use also had a higher mean daily frequency of SMBG, but no difference in bolus frequency was evident. Detailed analyses of alterations in insulin delivery and participant behaviours were, however, limited by the small numbers involved in our study. The key question of how individuals employ this technology to improve their glycaemic control therefore requires further enquiry. We acknowledge that the benefits seen in our study relate to short-term use of ‘patient-led’ sensor-guided pump management. Further RCTs of longer duration are required to determine the durability of these effects, as frequency of sensor use and behavioural responses to this technology may alter over time.

Recently, high user satisfaction and patient preference for RT-CGM were reported for a cohort with poor glycaemic control using insulin pump therapy [16]. Our findings demonstrate that, with basic education in the use of this technology, some patients with type 1 diabetes can indeed adapt to RT glycaemic information and successfully incorporate it into their diabetes management. Nevertheless, the relatively high rate of participant withdrawal and failure to comply with protocol-stipulated sensor use amongst a motivated patient population suggests that patient-led use of this technology is not universally acceptable and is too high an additional burden for some. In light of the additional expense associated with continuous use of RT-CGM, identification of those most likely to benefit from this technology will be an important focus of future studies.

In conclusion, this RCT has shown that patient-led use of sensor-guided pump management offers additional benefits for glycaemic control when compared with standard insulin pump therapy regimens. Improvement in HbA1c in such a well-controlled cohort, despite no additional patient–clinician contact over the study period, is clinically significant and has encouraging implications for future use of this technology. Ongoing technological improvements and the introduction of algorithms to guide responses to RT-CGM data are likely to optimise use of this management tool.