Introduction

The benefits of continuous glucose monitoring (CGM) in type 1 diabetes have been established among adults using insulin pumps. The DIAMOND randomized clinical trial examined the effectiveness of using CGM in improving glycemic control in participants using insulin injections. The frequency of hypoglycemic events in this trial has not been previously examined.

Methods

Adults with type 1 diabetes using multiple daily insulin injections (MDI) with A1C values of 7.5% to 9.9% and not using CGM were randomized to adopt CGM (CGM group, n = 105) or continue with usual care (control group, n = 53). CGM data were collected from both groups at the beginning of the study and after 3 and 6 months. A hypoglycemic event was defined as a series of at least CGM values less than 3.0 mmol/L, separated by 20 min or more, with no intervening values of 3.0 mmol/L or more. Hypoglycemic event rates per 24 h were compared using a linear model adjusted for the baseline event rate per 24 h, baseline A1C, and site as a random effect.

Results

In the CGM group, the median hypoglycemic event rate fell by 30% (0.23 per 24 h at baseline and 0.16 per 24 h at follow-up) while in the control group the rate was nearly unchanged (0.31 per 24 h at baseline and 0.30 per 24 h at follow-up; p value = 0.03).

Conclusion

In the DIAMOND randomized controlled trial, participants in the CGM group experienced a greater reduction in hypoglycemic event rate than participants receiving usual care in the control group.

Trial Registration

Clinicaltrials.gov Identifier: NCT02282397.
Literature
1.
Santiago JV. Lessons from the diabetes control and complications trial. Diabetes. 1993;42(11):1549–54.CrossRefPubMed
2.
Cariou B, Fontaine P, Eschwege E, et al. Frequency and predictors of confirmed hypoglycaemia in type 1 and insulin-treated type 2 diabetes mellitus patients in a real-life setting: results from the DIALOG study. Diabetes Metab. 2015;41(2):116–25.CrossRefPubMed
3.
Cryer PE. Preventing hypoglycaemia: what is the appropriate glucose alert value? Diabetologia. 2009;52(1):35–7.CrossRefPubMed
4.
Frier BM. Defining hypoglycaemia: what level has clinical relevance? Diabetologia. 2009;52(1):31–4.CrossRefPubMed
5.
International Hypoglycaemia Study Group. Glucose concentrations of less than 3.0 mmol/L (54 mg/dL) should be reported in clinical trials: a joint position statement of the American Diabetes Association and the European Association for the Study of Diabetes. Diabetes Care. 2017;40(1):155–7.CrossRef
6.
Beck RW, Riddlesworth T, Ruedy K, for the DIAMOND Study Group, et al. Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND Randomized Clinical Trial. JAMA. 2017;317(4):371–8.CrossRefPubMed
7.
Russell SJ, Beck RW. Design considerations for artificial pancreas pivotal studies. Diabetes Care. 2016;39(7):1161–7.CrossRefPubMed

Be confident that your patient care is up to date

Medicine Matters is being incorporated into Springer Medicine, our new medical education platform. 

Alongside the news coverage and expert commentary you have come to expect from Medicine Matters diabetes, Springer Medicine's complimentary membership also provides access to articles from renowned journals and a broad range of Continuing Medical Education programs. Create your free account »